55 | P a g e 
 
 
2.5.4 Presentation of Results 
2.5.4.1 The result of the calibration shall be presented in a standard form and will contain 
the following information: 
a) 
Prover serial number, type, dimensions and the proving location. 
b) 
Field Standard (test measure) – Nominal capacity identification and 
certifications. 
c) 
Observed values – pressure, temperature, density and volumes. 
d) 
Corrections for temperature, pressures, Prover materials and proving liquid. 
e) 
Base Prover Volume calculation. 
           
2.5.4.2 Results of calibration by master meter method shall be in the standard form giving 
information on the Prover, master meter and how its meter factor was established 
on site and the Base Prover volume calculation. 
 2.5.4.3 For Prover calibration using a Master meter, the master meter calibration flow rate 
shall agree with the flow rate at which the Prover is to be calibrated.  
2.5.4.4 The calibration result shall be endorsed by representative of Department of 
Petroleum Resources and approved by the Director of Petroleum Resources if found 
satisfactory and this shall constitute approval for the continued use of the Prover 
concerned. 
2.5.4.5 Input or change of Base Prover Volume and other parameters shall ONLY be in the 
presence of an officer of the Department of Petroleum Resources. 
2.5.4.6 For conduct of linearity test after a prover loop certification exercise, a newly 
generated BPV shall be used and immediately replaced with the old approved BPV 
pending the approval of the new BPV by the Department. 
 
2.6        Meter Proving Operations 
2.6.1     Definition 
This is the procedure used to determine the relationship between the true volume of a liquid 
measured by a meter and the volume indicated by the meter. This relationship is generally 
called the meter factor which is expressed as the ratio of the actual volume of a liquid 
passed through the meter to the volume registered by the meter. 
2.6.2   General Consideration 
2.6.2.1 Meters shall be proved as close to their normal operational conditions as possible.  
2.6.2.2 Meters shall be proved with liquid to be metered, or if not practicable such as the 
case of in-factory tests, with a liquid whose density and viscosity are very close to 
those of the liquid to be measured. 
2.6.2.3 Five consecutive proving runs must check to 0.020% repeatability, else the meter shall 
be taken out for inspection and repair before being put back into service.