65 | P a g e 
 
 
hydrometer/densitometer shall be used as the test liquid. The secondary standard 
hydrometer/densitometer to be used in the exercise shall have a valid certificate of its prior 
calibration against a primary standard hydrometer/densitometer which shall be issued by a 
recognized bureau of standard. 
2.2.6 Presentation of Results 
A report of the calibration exercise shall be submitted to the Department of Petroleum 
Resources for approval before the hydrometer/densitometer can be used. 
This report shall be accompanied by a copy of the certificate of the calibration of the 
secondary standard hydrometer/densitometer that has been used as reference during the 
exercise. 
2.3 Calibration of Pressure and Differential Pressure Measuring Devices 
2.3.1 Accreditation  
Only companies or Institutions that possesses a current Petroleum Resources registration 
permit as an equipment calibrator shall be eligible to carry out Pressure and Differential 
Pressure Measuring instrument calibration with participation of representative(s) of The 
Department of Petroleum Resources. 
2.3.2 Principle 
Pressure and Differential Pressure Measuring devices shall be calibrated by a pressure 
calibrator that is certified as intrinsically safe.  The Pressure Calibrator shall have the ability 
to apply pressure to a pressure transmitter or pressure switch while simultaneously 
measuring the mA signal or switch contacts.  
2.3.3 Frequency of Calibration 
All Pressure and Differential Pressure Measuring devices shall be calibrated and certified in 
the presence of DPR official before being put to use and thereafter annually.  
2.3.4 Applicable Standard 
Calibration method shall be as outlined in latest applicable ASTM and American Petroleum 
Institute Manual of Petroleum Measurement Standard (API MPMS). 
2.3.5 Verification and Calibration 
2.3.5.1 All new Pressure and Differential Pressure Measuring devices shall be deemed 
calibrated before use and annually thereafter. Certificate of calibration issued by the 
manufacturer shall be acceptable as evidence of calibration.  
2.3.5.2 If a Pressure and Differential Pressure Measuring device calibration error exceeds 
one-half of the scale division, error correction shall be applied whenever the 
instrument is used. 
 
2.3.6 Presentation of Results 
A report of the calibration exercise shall be submitted to the Department of Petroleum 
Resources for approval before the Pressure and Differential Pressure Measuring device can 
be used. 
This report shall be accompanied by a copy of the certificate of the calibration of the 
secondary standard Pressure and Differential Pressure Measuring device that has been used 
as reference during the exercise.