1.12.1 General Requirements  
1.12.1.1 Gas quality 
The meter shall as a minimum requirement, operate with any of the normal range natural 
gas composition mixtures specified in AGA Report No.8, this includes relative density 
between 0.554 (pure methane) and 0.87. 
1.12.1.2Temperature measurement 
Coriolis flow meters should operate over a temperature range of –40 to 200F (-40 to 93C). 
However, in extreme ambient environment, appropriate measures such as providing shade 
to reduce impact of direct sunlight to the meter should be considered. A thermowell should 
be installed upstream of the Coriolis sensor for temperature verification purposes  
1.12.1.3 Pressure measurement  
The location of the pressure measurement should be made near the sensor. The pressure 
tap can be located either upstream of the sensor.  
1.12.1.4 Vibration and Electrical noise 
Coriolis meter package should be installed far away from vibration sources like 
compressors, pumps, generators and electrical noise such as electromagnetic interference.  
1.12.1.5 Piping Configuration  
The following components should be provided in the piping design: Isolation block valves, 
Pressure, blow down port and thermowell port. The pipe-works must be aligned and 
properly supported to minimize stress due to pressure, weight and temperature. 
1.12.1.6 Mounting 
a. The sensor should be properly mounted, supported and aligned with the inlet and outlet 
flanges. A spool piece should be used in place of the meter to align pipe-work during the 
construction phase.    
b. The Coriolis transmitter shall be mounted to ensure easy access to communication 
equipment, displays and keypads. The Coriolis meter may have the transmitter mounted to 
the sensor or mounted remotely.  
1.12.2 Orientation  
The sensor tubes should be positioned in such a way that the possibility of heavier 
components, such as condensate, settling in the vibrating portion of the sensor is 
minimized.  
1.12.3 Minimum Performance Requirements 
The following minimum performance requirements shall be met by the meter, as supplied 
directly from the manufacturer, prior to making any calibration factor adjustments based 
on an independent third-party flow calibration of the meter. 
Repeatability:  
+ 0.35% of reading for Qt < Qi < Qmax 
 + 1.0% of reading for Qmin < Qi < Qt 
  
Where  
Qt is the transitional mass flow rate.